Accuracy vs. Precision: Why Your Instrument Can Be One Without the Other
Understand the difference between precision and accuracy, with a 100.000 mm calibration example and how to read as-found and as-left certificate data.
The difference between precision and accuracy is what each one measures. Accuracy is how close a measured value sits to the accepted reference value. Precision is how closely repeated measurements of the same thing agree with each other. Because they describe separate properties, an instrument can be precise but inaccurate, or accurate on average but imprecise.
One tells you whether your readings land on the right answer. The other tells you whether they land in the same place twice. Knowing which you are looking at turns a page of calibration data into a decision you can defend.
Why people confuse accuracy and precision
In everyday speech, the two are interchangeable. Someone calls a scale precise when they mean it reads correctly, or a caliper accurate when they mean it repeats well.
Measurement science separates them because they point to different faults. Poor precision means random variation: technique, environment, play in a mechanism, thermal drift during a run. Poor accuracy means a systematic offset such as a wrong zero, a shifted span, or wear on a measuring face. Those get fixed in different ways, so collapsing the words hides the diagnosis.
The metrology nuance
Measurement accuracy is not a number. Under the International Vocabulary of Metrology (VIM), maintained by the Joint Committee for Guides in Metrology at the BIPM, measurement accuracy is not a quantity and is not given a numerical quantity value.
A measurement is more accurate when it has a smaller measurement error. So the "±0.02 mm accuracy" on a datasheet is a specification limit, not a measured accuracy figure for the instrument in front of you.
VIM gives a separate term for the part people usually mean when they quote a number. Measurement trueness is the closeness of agreement between the average of an infinite number of replicate measured values and a reference value, so it concerns systematic offset of the average.
Measurement precision is the closeness of agreement between replicate measurements under specified conditions, usually expressed numerically as a standard deviation.
Plain version: precision is scatter, trueness is where the middle of that scatter sits, and accuracy is the overall closeness both contribute to.
The four combinations of accuracy and precision
Every set of repeated measurements falls into one of four states. Two properties define which one: where the readings sit relative to the accepted reference value, which is accuracy, and how closely they agree with each other, which is precision, or repeatability in practice.
Accurate and precise
Readings sit close to the reference value and close to each other. Both the systematic and random components of error are small at that measurement point. This is the outcome you want before an instrument returns to service.
Precise but inaccurate
Readings agree closely with each other but sit consistently to one side of the reference value. Repeatability is good and the offset is stable. Likely causes include a shifted zero point, a span or gain error, mechanical wear, or drift since the last calibration.
This is the dangerous case: five identical readings look like confirmation when they are five copies of the same error.
Accurate on average but imprecise
Readings fall on both sides of the reference value and vary widely, so the average lands close to it. The meaning flatters the instrument. Any single reading, which is what an operator records on the shop floor, could be well off. Look for loose mechanisms, an unstable environment, or inconsistent technique.
Neither accurate nor precise
Readings vary widely and also sit off to one side. Troubleshooting has to address both sources at once: correcting the offset leaves the scatter, and steadying the scatter leaves readings that are still in the wrong place.
A calibration example: consistent readings can still be wrong
Calibration is fundamentally a comparison. Under VIM, calibration establishes a relation between the values provided by measurement standards, with their uncertainties, and the corresponding indications of the instrument under test. Adjustment may follow, but adjustment is a separate operation, and after one, the instrument normally has to be calibrated again.
The figures below are illustrative, not an acceptance criterion to copy. A dimensional measuring tool is checked against a 100.000 mm reference standard.
Read the as-found result
Five as-found readings before any adjustment:
- 99.921 mm
- 99.920 mm
- 99.922 mm
- 99.921 mm
- 99.921 mm
What the numbers say:
- Mean result: 99.921 mm
- Error versus the reference: 0.079 mm low
- Spread across five readings: 0.002 mm
Read the two properties separately. The spread is tight, so repeatability at this point is strong. The mean sits 0.079 mm below the reference, so there is a clear negative systematic error. This is the second case above in numbers: precise, and wrong.
Repeatability alone establishes nothing about acceptable measurement performance. An instrument that repeats to 0.002 mm while reading 0.079 mm low hands you the same wrong dimension every time.
What adjustment and as-left verification change
After adjustment and verification, five as-left readings:
- 100.000 mm
- 99.999 mm
- 100.001 mm
- 100.000 mm
- 100.000 mm
The mean is now 100.000 mm and the error at this point is effectively zero. Notice what did not change: the spread is still 0.002 mm. Adjustment moved the group onto the reference without tightening it, so a certificate showing an improved mean tells you nothing about improved repeatability.
Calibration records document what was observed, and the as-found result remains the evidence covering the period the instrument was already in use.
Why tolerance and uncertainty still matter
An error of 0.079 mm is not automatically a failure, and an error of 0.000 mm is not automatically a pass. Three things decide that:
- The tolerance required for the measurements the instrument is used for
- The measurement uncertainty reported with the calibration result
- The decision rule the laboratory applied when comparing the result and its uncertainty against that tolerance
Decision rules are what stop pass and fail from being a coin flip near the limit. ILAC publishes guidance on decision rules and statements of conformity in ILAC G8:09/2019, revised to support the conformity statement requirements in the 2017 edition of ISO/IEC 17025.
Some approaches apply a guard band, narrowing the acceptance zone by the uncertainty so a borderline result is not accepted on optimism. The rule used changes the verdict on identical numbers, which is why a certificate should state it.
For how uncertainty is built up and reported, see Tripoint's guide to temperature calibration, stability, uncertainty and traceability.
Accuracy, precision, and the terms they get confused with
Four pairs cause most of the trouble.
Precision vs. repeatability vs. reproducibility
Precision always applies under specified conditions, and the conditions are the point.
- Repeatability is precision under repeatability conditions: same measurement procedure, operators, measuring system, operating conditions and location, with replicate measurements over a short period.
- Reproducibility is precision under reproducibility conditions, a set that includes different locations, operators and measuring systems.
An instrument can repeat beautifully for one technician on one bench and scatter badly across three technicians in two rooms. When a specification claims precision, check which conditions it covers.
Accuracy vs. trueness
Trueness concerns only the closeness of the average result to a reference value. It is inversely related to systematic error and unrelated to random error, while accuracy covers both. This is why careful technical documents state bias, error, and uncertainty at specific measurement points instead of calling an instrument "accurate" outright.
Performance is never one fixed property of a device. It depends on the measurement point, range, method, environment, tolerance, reported uncertainty, and repeatability conditions.
Precision vs. resolution
Resolution is the smallest change in the measured quantity that causes a perceptible change in the indication. It is a display or detection property, and extra decimal places prove nothing. A digital indicator can show five digits, repeat poorly, and read 0.08 mm low all at once.
Resolution sets the finest increment you can observe. Precision tells you whether observing it again gives the same answer.
Accuracy and precision vs. measurement uncertainty
Uncertainty is the quantified doubt attached to a measurement result: a non-negative parameter characterising the dispersion of the quantity values being attributed to a measurand, based on the information used. It is not a synonym for error. Error is the gap between an indication and the reference value.
Uncertainty is the interval around a reported result, built from repeatability, the reference standard's own uncertainty, resolution and environmental effects. A result with no uncertainty statement cannot be compared to a tolerance defensibly.
What an as-found and as-left certificate can tell you
As-found data
As-found is the instrument's measured condition on arrival, before any adjustment or repair. It is the only evidence about how the instrument behaved while in service, which is why a failed as-found result is not just a maintenance event.
It usually triggers an impact assessment: which parts, batches or process records were accepted using that instrument since its last calibration, and whether that work needs review.
As-left data
As-left is the verified condition after any adjustment, repair or calibration work is finished. Identical as-found and as-left values are common and unremarkable, and usually mean the instrument was within its limits and no adjustment was needed.
What to look for beyond a pass or fail label
Read the data, not the verdict:
- Instrument identification and calibration date, matching the asset in hand
- Test points and actual readings, including as-found and as-left values
- The reference standard used and the traceability statement supporting it
- The reported measurement uncertainty at each point or range
- The applicable tolerance and any conformity statement, with the decision rule behind it
NIST is explicit that metrological traceability is a property of a measurement result, not of an instrument or a laboratory, and that it requires a documented unbroken chain of calibrations with uncertainty determined at each step.
Holding a certificate does not make every future reading traceable. The sticker on the housing is an identification aid showing what the asset is and when it is next due. The certificate holds the documented result, a split covered in calibration stickers, labels, dates and certificates.
When the difference affects real work
Dimensional inspection
A caliper with a consistent zero offset repeats to a few microns and reports every dimension low. Inspection records look clean. Parts get accepted that should have been rejected, or scrapped when they were fine, and the pattern surfaces at the next calibration or when a customer finds it.
Teams responsible for gauges, micrometers and calipers can review scope and test points with dimensional calibration services.
Weighing and process control
A scale can produce stable, repeatable readings that are consistently offset by a fixed amount or by a proportion of the load. Stability is reassuring and says nothing about trueness. Corner load, drift and zero handling each leave their own signature in the data, patterns covered in detail in Tripoint's guide to weighing scale calibration errors.
Temperature and electronic measurements
Here the conditions of measurement do most of the work. Stability at the test point, immersion depth, ambient conditions, drift between calibrations, and how many points across the range were checked all shape how a result reads.
An instrument verified at one point tells you little about the far end of its range, so test point coverage belongs in the scope discussion.
Use the right evidence for the measurement decision
The difference between precision and accuracy comes down to three separate questions. Closeness to the reference is accuracy. Consistency across repeated readings is precision. How much doubt surrounds the figure is uncertainty. Confusing them means fixing scatter when the problem is offset, or trusting a mean that hides unusable readings.
Measured data therefore outranks the word "accurate" on a datasheet. Test points, as-found and as-left values, the reference standard, the uncertainty, and the decision rule are what let you judge an instrument against the tolerance you work to.