    AMS2750 Pyrometry Guide: Requirements, Latest Revision (AMS2750H) and Audit-Ready Programs

    AMS2750 pyrometry explained: what the spec governs, furnace classes, instrumentation types, revision changes and how to build a Nadcap audit-ready program.

    AMS2750, also written AMS 2750, is a pyrometry specification for thermal processing equipment used in metal heat treatment. It sets requirements for temperature sensors, instruments, and the tests that confirm a furnace can produce and measure heat accurately. Two of those requirements, the temperature uniformity survey and the system accuracy test, sit at the center of almost every pyrometry compliance program.

    An AMS2750 program rests on two recurring furnace checks: a temperature uniformity survey that proves the work zone heats evenly, and a system accuracy test that proves the installed instruments read correctly. This guide covers how AMS2750 schedules, documents and audits both as one program.

    What AMS2750 Actually Governs

    AMS2750 is a pyrometry specification published by SAE International, the standards body that also maintains companion process specifications such as AMS2759 and AMS2770. The specification covers five technical areas: temperature sensors, instrumentation, thermal processing equipment classification, system accuracy tests, and temperature uniformity surveys. 

    SAE has revised the standard eight times since it was first issued in April 1980, most recently as AMS2750H in July 2024. The current revision, AMS2750H, was issued in July 2024 and supersedes AMS2750G, AMS2750F, and AMS2750E.

    AMS2750 does not set the process temperature for any specific material or part. That belongs to the process specifications that call it out, such as AMS2759 for general heat treatment of steel or AMS2770 for aluminum alloys. AMS2750 covers only the pyrometry infrastructure underneath those processes: the sensors, instruments, tests, and records that prove the equipment can hit and measure the required temperature.

    What Is Pyrometry?

    Pyrometry is the measurement of temperature in thermal processing, and in AMS2750 it means the full measurement system around a furnace: the sensors, the instruments they feed, the calibrations behind both, and the tests that prove the system reads and holds temperature correctly. When an auditor reviews your pyrometry, they are reviewing that whole chain, not a single thermocouple or controller.

    Nadcap, the accreditation program administered by the Performance Review Institute on behalf of subscribing aerospace OEMs, audits heat-treating suppliers against AMS2750 when a customer requires that accreditation. 

    AMS2750 compliance and Nadcap accreditation are related but not identical. A facility can meet AMS2750 requirements without holding Nadcap accreditation, and the exact audit criteria a Nadcap auditor applies depend on the customer specifications that flow down to the supplier. Confirm which framework your own customers require rather than assuming Nadcap governs every AMS2750 activity you perform.

    AMS2750 Latest Revision: What Changed in AMS2750H

    The latest revision of AMS2750 is AMS2750H, revised by SAE in July 2024. It replaced the AMS2750G, issued in June 2022. AMS2750F was issued in June 2020 and AMS2750E in July 2012. Each revision has tightened or clarified requirements rather than replacing the structure of the standard, so most programs built for Rev G carry forward, but several day-to-day rules changed.

    • AMS2750F introduced a rule that still governs interval extensions today. The initial TUS performed on a new or modified furnace does not count toward the consecutive passing surveys required to extend the testing interval. That count starts only with the first periodic survey after the initial one. 
    • AMS2750F also tightened the accreditation requirement for third-party pyrometry providers, requiring ISO/IEC 17025 accreditation with a scope that specifically references AMS2750, not a general laboratory accreditation with no stated scope.
    • AMS2750H clarified which sensors count as expendable: braided fiberglass or plastic-insulated sensors are treated as expendable even inside a closed protection tube, while ceramic-insulated bare-wire sensors in a closed tube are nonexpendable.
    • AMS2750H tightened sensor records. Batch numbers must be recorded for single-use sensors, load sensor batches must be traceable to the load, and nonexpendable TUS sensors are capped at 270 uses.
    • AMS2750H prohibits extrapolating correction factors beyond the calibrated points and requires any interpolation method to be defined and applied consistently.
    • AMS2750H changed TUS data collection: recording starts at least 100 degrees F (55 degrees C) below each survey temperature, readings are recorded simultaneously, and hot and cold recording locations are assessed individually.
    • AMS2750H states that production cannot start once interval extension days have expired, and that third-party provider accreditation requirements apply to on-site work.

    Exactly which clauses changed between AMS2750G and AMS2750H, and how those changes apply to a specific furnace class and instrumentation type, requires comparing the two controlled texts directly. Treat any revision summary, including this one, as a starting point for that comparison rather than a substitute for it.

    TUS vs SAT: What Does Each Test Verify?

    A temperature uniformity survey and a system accuracy test check two different failure modes in the same furnace. AMS2750 requires both because passing one test says nothing about the other.

    • Core purpose: TUS confirms the furnace heats evenly across its qualified working zone. SAT confirms the installed measurement system reads that temperature accurately.
    • What it measures: TUS maps temperature at multiple sensor locations across a defined zone during a soak period. SAT compares one measurement point against a calibrated reference instrument.
    • What a failure means: A failed TUS shrinks or voids the qualified working zone until a passing survey is on file. A failed SAT triggers a review of everything processed since the last passing test.
    • What it does not replace: A passing TUS is not a substitute for SAT, and a passing SAT is not a substitute for TUS. AMS2750 treats them as separate, mandatory requirements.

    For the full mechanics of each test, see our temperature uniformity survey basics guide and our system accuracy test guide.

    AMS2750 allows a standard SAT, an alternate SAT for sensors that are replaced often, and a documented exemption; which one applies, and how to calculate the result, is covered in our system accuracy test procedure guide. A TUS qualifies a defined working zone, and parts must be processed inside that zone at the operating range the survey covered. Our temperature uniformity survey guide explains how the qualified working zone is defined and documented in the report.

    AMS2750 Furnace Classes and Instrumentation Types

    AMS2750 sorts furnaces into six classes based on how tightly they must hold temperature across the working zone, and separately into instrumentation types based on which sensors and safeguards are wired into the control loop. Class and instrumentation type describe two different things about a furnace, and neither one substitutes for the other.

    The six furnace classes, by allowable temperature variation across the working zone, are:

    1. Class 1: ±5°F (±3°C)
    2. Class 2: ±10°F (±6°C)
    3. Class 3: ±15°F (±8°C)
    4. Class 4: ±20°F (±10°C)
    5. Class 5: ±25°F (±14°C)
    6. Class 6: ±50°F (±28°C)

    These uniformity values are corroborated byEurotherm's AMS2750 pyrometry reference; confirm the exact figures against your governing revision before using them to qualify equipment.

    Instrumentation type describes what the control loop actually includes:

    • Type A: control, recording, hot and cold zone monitoring, load monitoring, additional recording, and over-temperature protection.
    • Type B: control, recording, hot and cold zone monitoring, load monitoring, and over-temperature protection.
    • Type C: control, recording, hot and cold zone monitoring, and load monitoring.
    • Type D+: control, recording, and hot and cold zone monitoring.
    • Type D: control and recording only.
    • Type E: control only.

    These are uniformity and instrumentation categories, not a complete restatement of AMS2750's requirements for each one. The governing revision defines exact tolerances, sensor placement, and documentation for every class and type combination, and a furnace's class does not tell you its instrumentation type or vice versa. Confirm both before scheduling TUS and SAT work, since the required frequency for each test depends on the combination of the two, not on class alone.

    AMS2750 Calibration Planning: TUS, SAT and Instrument Schedules

    TUS frequency depends on furnace class and instrumentation type together, not on class alone. A furnace with a tighter class and more complete instrumentation may start on a different initial interval than a furnace with a looser class and simpler instrumentation, and the interval can extend only after the applicable number of consecutive passing surveys, counted from the first periodic survey rather than the initial one. 

    • Confirm the governing table in your current revision before assuming one pass count applies across an entire fleet of furnaces.
    • A furnace also needs a new TUS any time it undergoes a modification, repair, or change that could affect its thermal performance, and a furnace qualified across multiple operating ranges needs a TUS in each range it runs.
    • SAT frequency follows the same logic: it depends on which SAT method applies to a given sensor and furnace configuration, not on a single blanket schedule. Every instrument in the measurement loop, meaning the controller, recorder, and load thermocouples, needs its own calibration schedule tied to furnace class and instrumentation type. 
    • Calibration points must cover the instrument's operating range at intervals defined by the applicable specification and instrument category. AMS2750 does not apply one universal spacing rule to every sensor and instrument combination, so check the table that applies to your specific equipment rather than a rule of thumb.
    • Traceability for those calibrations must run back through an unbroken chain to the relevant national metrology institute, most often the National Institute of Standards and Technology in the United States, though the applicable specification and your customer's requirements determine which traceability routes are acceptable. Document the actual chain your provider uses instead of assuming one route is the only path.

    Set internal due dates at least two weeks ahead of the AMS2750-required date for every test. That buffer leaves room to reschedule or rerun a test if a provider has a conflict or a survey comes back marginal, without missing the actual compliance date.

    Nadcap Heat-Treating Audit Preparation

    Nadcap audits for heat-treating pyrometry follow a checklist administered by PRI that covers equipment configuration, calibration status, TUS and SAT records, and corrective action history. That checklist is revised periodically, so confirm you are working from the current version and any customer-specific supplements before an audit window opens, rather than relying on a summary from a prior audit cycle.

    A few findings show up repeatedly in pyrometry audits. Third-party providers get cited when their ISO/IEC 17025 accreditation scope does not specifically list AMS2750 and the exact activities performed; a general accreditation with no stated scope is not enough. 

    Thermocouple location diagrams get flagged when they omit the control thermocouple position or mislabel sensor numbers relative to the TUS report. Suppliers get cited when they cannot show they reviewed and accepted a third-party TUS or SAT report internally, even when the report itself is correct. That internal review is a separate, documented requirement from the test itself.

    Build your preparation around the same categories an auditor checks: which specification and customer supplements apply to each furnace, whether the equipment record matches what is actually installed, whether calibration certificates cover the right methods and ranges, whether your schedule and any interval extensions are justified by the applicable conditions, and whether every TUS and SAT report has a documented internal review and disposition. 

    Our Nadcap pyrometry calibration readiness guide walks through these record categories in more detail.

    What to Do When Results Fail

    A failed SAT should trigger your documented nonconformance process, not a single default reaction. Identify the affected measurement system, restrict further use of it as required, and evaluate what was processed since the last passing test. Some of that product may need containment, re-inspection, or reprocessing. Some may not. 

    • The right disposition depends on the process specification, the customer's requirements, and how far the reading fell outside tolerance, so document the investigation and the decision rather than assuming every failure ends the same way.
    • A failed TUS reverts the qualified working zone to an unqualified state, and production within that operating range stops until a passing survey is back on file. Because the qualified zone can be smaller than the furnace's physical interior, confirm exactly which zone and operating range failed before assuming the entire furnace is out of service.
    • Keep both TUS and SAT records current, signed off by a qualified reviewer, and organized so a Nadcap auditor or customer reviewer can trace the history for any furnace on short notice.

    Selecting an AMS2750 TUS/SAT Service Provider

    Start with your equipment and requirements, not a vendor list. Identify the furnace type, operating range, qualified working zone, instrumentation type, and the AMS2750 revision your customer requires before you request a quote.

    Does AMS2750 Require ISO/IEC 17025 Accreditation?

    Evaluate the accreditation requirement for the specific activity and provider arrangement under the applicable AMS2750 revision and your customer's requirements. A provider's general accreditation claim is not enough on its own. 

    Ask any provider to confirm their ISO/IEC 17025 accreditation scope by name, specifically whether it lists AMS2750 and the exact activity you need, such as TUS, SAT, or a given instrument category, before assuming coverage.

    Tripoint lists TUS, SAT, and temperature-instrument calibration among its temperature calibration capabilities. Its current scope of accreditation documents which specific activities, ranges, and locations are covered. Review that scope alongside your equipment list before assuming any single service is included by default. A provider's accreditation can change at renewal, so confirm coverage at the time of your engagement rather than from memory.

    Ask about turnaround, report contents, and how the provider documents the TUS working zone and the SAT difference calculation. A report that omits the qualified zone boundaries or the correction factor treatment will not hold up under a Nadcap auditor's review, no matter how well the test itself was performed.

    AMS2750 FAQ

    What is AMS 2750?

    AMS2750 is the SAE International pyrometry specification for thermal processing equipment. It sets requirements for temperature sensors, instruments, furnace classification, system accuracy tests and temperature uniformity surveys used in heat treating, mostly in aerospace supply chains.

    What is the latest revision of AMS 2750?

    The latest revision is AMS2750H, revised by SAE in July 2024. It replaced AMS2750G from June 2022.

    What is the difference between AMS2750 compliance and Nadcap accreditation?

    AMS2750 is the specification. Nadcap is an industry accreditation program, run by the Performance Review Institute, that audits suppliers against AMS2750 and related checklists when their customers require it. A facility can meet AMS2750 without holding Nadcap accreditation.

    What does an AMS2750 system accuracy test check?

    A system accuracy test checks whether the installed sensor and instrument read the furnace temperature correctly by comparing them with a calibrated test sensor and instrument at the same point. Our system accuracy test guide walks through the full procedure.

    Building an Audit-Ready Pyrometry Program

    AMS2750 pyrometry compliance comes down to two tests that answer different questions, a class and instrumentation system that determines how often each test runs, and documentation detailed enough to survive an audit built around all of it. None of that works as a set of boxes checked once a year and forgotten.

    • Assign each furnace a schedule tied to its actual class and instrumentation type, verify provider accreditation scope before every engagement rather than assuming it carries over, and review every TUS and SAT report internally before filing it. Track consecutive passing surveys from the first periodic TUS forward, not from the initial one, before approving any interval extension.
    • If your TUS and SAT schedule needs a second look, or you are not sure which furnace class and instrumentation combination applies to your equipment, request a quote with your furnace list, operating ranges, and required test dates, or contact our team to walk through your specific AMS2750 requirements before your next audit window opens.

    Published April 15, 2026. This guide was last reviewed on October 14, 2026, against SAE's published AMS2750 revision record, PRI's public Nadcap program documentation, and Eurotherm's AMS2750H pyrometry reference for furnace class and instrumentation data. 

    For clause-specific interpretation of your own equipment, confirm current requirements against the controlled AMS2750H text with your pyrometry lead or your Nadcap auditor.

    Related Resources

    Temperature Calibration
    Apr 15, 2026

    Temperature Bath Calibration: Stability, Uniformity, and Best Practices

    Temperature bath calibration hinges on stability, uniformity, and immersion depth. Learn NIST-traceable best practices and ISO/IEC 17025 requirements for accurate results.

    Temperature Calibration
    Mar 19, 2026

    A Complete Guide to Chart Recorder Calibration: Channels, Scaling, Alarms, and Compliance Records

    Temperature Calibration
    Mar 19, 2026

    Temperature Controller Calibration: Procedure, Records, and Tuning

    See how temperature controller calibration works: the full procedure, what a calibration certificate should include, and how it differs from PID tuning.