System Accuracy Test (SAT): Step-by-Step Procedure Guide
A system accuracy test compares a furnace sensor to a calibrated test system. See the AMS2750 SAT procedure, SAT vs TUS, & the records auditors want.
A system accuracy test, or SAT, checks whether a furnace temperature measurement system is reading the true temperature inside the equipment. It compares the installed control sensor and instrument against an independent, calibrated test system, then documents the difference between the two readings. The test is one of the core pyrometry checks that keeps a furnace qualified for heat treatment work.
The test answers a single question: can the permanent thermocouple and controller be trusted at the setpoint you run? A SAT does not check temperature uniformity across the work zone. It checks accuracy at one location, for one measurement channel, at one or more operating temperatures. That focus is what makes the procedure fast to run and easy to document when you know the steps.
This guide walks through what a SAT verifies, when it is required, how a comparison SAT differs from an alternate SAT, the equipment you need, and a step-by-step procedure. It also covers the records auditors expect, the failures that trip teams up most often, and what to do when a test does not pass. The aim is a page you can hand to a new technician or use to prepare for an audit.
What Is a System Accuracy Test?
A system accuracy test measures the error in a furnace temperature measurement channel. The channel includes the control or monitoring sensor, the extension wire and connectors, and the instrument that displays or records the reading. The SAT proves that the combined error of all those parts stays within the tolerance your furnace class allows.
The standard SAE AMS2750 defines the SAT as an evaluation of the combined errors, or the sum of correction factors, of the sensor, the extension cable and connectors, and the instrument. In plain terms, every part of the measurement path adds a small error. The SAT adds those errors together and confirms the total is acceptable.
A SAT is a point check, not a survey. One test covers one sensor location and the channel attached to it. Furnaces with multiple control zones need a SAT on each zone. The result is a number: the difference between what the furnace says and what the independent test system says, after correction factors are applied.
What an AMS2750 SAT Verifies
An AMS2750 SAT verifies the accuracy of the whole control or recording channel, not just the thermocouple. Each link in the chain can drift, and the test catches the combined effect.
The channel a SAT checks includes several parts:
- The control sensor or zone thermocouple installed in the furnace
- The extension wire that carries the signal back to the instrument
- The connectors and terminal blocks along that path
- The controller, recorder, or high-limit instrument that reads the signal
- The correction factors applied to compensate for known calibration error
Correction factors are central to the test. Every calibrated sensor and instrument carries a certificate that lists its error at specific temperatures. During a SAT, those correction factors are pulled from the certificates and applied to the readings so the comparison reflects true temperature, not raw display values. When a test temperature falls between two calibrated points, AMS2750F and later revisions allow you to use the nearest correction or to interpolate between the two closest values.
The reason the test covers the full chain matters for compliance. A thermocouple can be perfectly calibrated while a worn connector or a miswired extension lead introduces several degrees of error. A SAT that only checked the sensor would miss that. By comparing the complete installed system against an independent reference, the test exposes error wherever it lives.
When a SAT Is Required
A SAT is required on a recurring schedule set by your furnace class and instrumentation type, and again whenever the measurement system changes. The exact interval lives in the AMS2750 frequency tables, so always work from the current revision rather than memory.
Two triggers drive testing. The first is the calendar. Furnaces run on a fixed SAT frequency, often quarterly for many classes, with tighter intervals for more demanding furnace classes. The second trigger is change. Replacing a control thermocouple, swapping an instrument, or altering the extension wiring can all require a fresh SAT before the furnace returns to qualified production.
The 2024 transition guidance from heat-treatment specialists notes that newer AMS2750 revisions added flexibility around SAT frequency, including clearer rules for waivers and alternate methods that reflect the rise of advanced monitoring. Even with that flexibility, the principle holds: the measurement system must be proven accurate before you trust it for parts. For planning intervals and aligning them with calibration cycles, map your schedule against the requirements in your AMS2750 calibration planning workflow.
Comparison SAT vs Alternate SAT
The two main SAT methods are the comparison SAT and the alternate SAT. They reach the same goal, proving channel accuracy, but they get there differently and apply to different sensor situations.
A comparison SAT is a physical, in-furnace test. You place a calibrated test thermocouple close to the furnace control sensor, read both systems at temperature, apply correction factors to the test side, and record the difference. The standard defines this method as a comparison of the difference between the readings of the equipment sensor system under test and the corrected reading of the test sensor system. It is sometimes called a probe SAT, and it is the method most teams picture when they think of a SAT.
An alternate SAT is a mathematical method rather than a live probe test. It sums the known errors and correction factors of the furnace sensor with the calibration error of the connector, extension cable, and instrument. The alternate SAT exists for situations where a comparison test is impractical, most often when expendable thermocouples are replaced every cycle or far more frequently than the standard SAT interval. If you use nonexpendable thermocouples and replace them on the normal schedule, you typically run a standard comparison SAT. If you swap expendable sensors each use, the alternate SAT path usually applies, and instrument calibration must take place at the location the standard specifies within the sensor system.
Choosing the right method comes down to how you manage sensors. Match the method to your thermocouple practice, document why you chose it, and keep that reasoning with your records. Auditors look for a defensible rationale, not just a passing number.
SAT Equipment and Setup
A comparison SAT uses two pieces of independent equipment: a calibrated SAT thermocouple and a field test instrument. Together they form the reference system that the installed furnace channel is measured against.
The SAT thermocouple is the calibrated test sensor placed near the furnace control sensor. It carries its own calibration certificate and correction factors. The field test instrument is the portable, calibrated readout that displays the SAT thermocouple's temperature. Industry guidance from instrumentation makers describes the SAT as a check performed with an independent device system, a field test instrument paired with a calibrated SAT thermocouple, where the delta between test and control readings shows the ongoing accuracy of the installed setup.
Sensor proximity is the setup detail that decides whether a test is valid. The test thermocouple must sit close to the furnace control sensor so both measure the same thermal conditions. Pyrometry guidance puts the AMS2750 limit at 7.5 cm between the control sensor and the test sensor, while the automotive CQI-9 standard holds a tighter 5 cm. Place the sensor too far away and the reading reflects a temperature gradient, not channel error.
Good setup also means letting the system settle. Bring the furnace to the chosen setpoint, allow temperatures to stabilize, and only then capture readings. Run the test at the temperatures your process actually uses, since accuracy can vary across a furnace's range.
Step-by-Step System Accuracy Test Procedure
Performing a system accuracy test follows a repeatable sequence: check records, place the test sensor, capture and correct readings, then document the result. Each step protects the validity of the next.
- Review equipment and calibration records. Confirm the SAT thermocouple and field test instrument both hold current calibration certificates. Pull the correction factors for the temperatures you plan to test. Verify the furnace's installed sensor and instrument are within their own calibration windows. A test run with expired calibration is not valid no matter how clean the numbers look.
- Place the SAT thermocouple. Position the calibrated test sensor within the required proximity of the furnace control sensor, no more than 7.5 cm under AMS2750. Make sure the test sensor tip and the control sensor sit in the same thermal zone so both see the same conditions.
- Stabilize at the setpoint. Bring the furnace to the chosen operating temperature and let it settle. Readings taken while the furnace is still trending toward setpoint will not reflect true channel accuracy.
- Capture and correct readings. Record the furnace system reading and the field test instrument reading at the same moment. Apply the correction factors from the certificates to the test system. When the test temperature falls between calibrated points, use the nearest correction or interpolate per the standard.
- Calculate the deviation. Subtract the corrected test reading from the furnace system reading. This difference is the system accuracy error for that channel at that temperature.
- Compare against the tolerance. Check the calculated deviation against the allowable SAT tolerance for your furnace class and instrumentation type, as listed in the AMS2750 tables. A result inside tolerance passes. A result outside it fails and triggers corrective action.
- Document the pass or fail result. Record every input and output so the test stands up to audit. That documentation is where most of the compliance value sits.
Run this sequence for each control zone that requires a SAT. Keep the conditions consistent from test to test so trends in the data mean something over time.
How a SAT Differs From a TUS
A SAT verifies one measurement channel at one point. A temperature uniformity survey verifies that temperature is even across the whole qualified work zone, and the two run on separate AMS2750 frequency tables. For a side-by-side comparison, see the TUS vs SAT section of our AMS2750 guide.
Common SAT Failures and How to Prevent Them
Most failed or invalid SATs trace back to a small set of avoidable causes. Knowing them ahead of time keeps a test from being thrown out after the furnace has already been tied up.
The failures that come up most often include:
- Expired thermocouple or instrument calibration. A test sensor or field instrument past its certificate date invalidates the entire test. Check dates before you start.
- Incorrect or missing correction factors. Using raw readings, the wrong certificate, or a correction for the wrong temperature corrupts the deviation calculation.
- Sensor placement beyond the proximity limit. A test thermocouple sitting too far from the control sensor measures a gradient, not channel error.
- Recorder and controller mismatch. Reading the wrong instrument channel, or one that is not the qualified control or monitoring instrument, produces a result that does not represent the system.
- Damaged or miswired extension wire. A degraded lead, a reversed polarity connection, or a worn connector adds error that the test will flag, sometimes after a lot of confusion about the source.
- Missing or incomplete records. A passing number with no supporting documentation will not survive an audit.
Prevention is mostly discipline. Verify calibration first, confirm proximity, read the correct channel, and capture every value as you go rather than reconstructing it later.
AMS2750 SAT Records Checklist
Audit-ready SAT records capture the full story of the test, from the equipment used to the final pass or fail decision. A passing result without documentation does not demonstrate compliance.
A complete SAT record set should include:
- Furnace identification and the specific zone or sensor tested
- The setpoint or test temperature, and the operating range it supports
- The installed furnace sensor and instrument reading
- The SAT thermocouple reading and the field test instrument reading
- Calibration certificate references for the SAT thermocouple and field test instrument
- The correction factors applied and the source certificate for each
- The calculated deviation and the allowable tolerance it was compared against
- The pass or fail result
- The technician name, the date, and the equipment identifiers used
Keep these records traceable and tamper-resistant. Recent AMS2750 revisions tightened electronic record-keeping expectations, calling for digital records that are protected and aligned with modern data acquisition systems. The cleaner your records, the faster an audit goes, which is why SAT documentation sits at the center of Nadcap pyrometry readiness.
What to Do If a SAT Fails
A failed SAT means the measurement channel read outside tolerance, so the furnace cannot run qualified work on that zone until the cause is found and fixed. Treat a failure as a stop-and-investigate event, not a number to retest until it passes.
Start with an impact review. Determine which production ran since the last passing SAT, because parts processed on an inaccurate channel may need engineering review. Then troubleshoot the chain in order: confirm the test equipment calibration was valid, check correction factors, inspect the extension wire and connectors, and verify you read the correct instrument. Many failures resolve at one of these points rather than at the sensor itself.
Once you identify and correct the cause, run the SAT again and document the corrective action alongside the new result. Keep both the failed and the passing records. The corrective-action trail is exactly what an auditor wants to see, and it shows the system is controlled rather than merely lucky.
Get Your Furnaces SAT-Ready
A SAT is only as good as the calibrated equipment and records behind it, and most failed or rejected tests trace back to one of those two gaps.
Our temperature calibration capabilities cover the SAT thermocouples, field test instruments, and correction-factor certificates the procedure depends on, while our AMS2750 audit-ready pyrometry program keeps your test schedule, documentation, and Nadcap pyrometry readiness aligned across every furnace.